- Precision.
- Specificity.
- Limit of detection.
- Limit of quantitation.
- Linearity.
- Range.
- Robustness.
Checklist for Analytical Method Validation (Chemical)
Assay/Related Substances
TEST |
ASSAY/RELATED
SUBSTANCES |
||
PARAMETER |
No. |
DOCUMENTS REQUIRED |
AVAILABILITY |
Specificity |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Chromatogram/spectrum
for the following solutions: |
|
|
Standard |
|
||
Sample |
|
||
Blank/Placebo |
|
||
Spike
solution |
|
||
System
suitability tests |
|
||
4 |
Impurities
available |
|
|
Peak
purity for PDA detector or spike solution for non-PDA detector |
|
||
Relative
Retention Time (RRT) |
|
||
Resolution
(R) where applicable |
|
||
5 |
Impurities
not available |
|
|
Stress
data (minimum data for humidity, heat and light) |
|
||
Peak
purity |
|
||
Relative
Retention Time (RRT) |
|
||
Resolution
(R) where applicable |
|
||
Linearity |
1 |
Testing
Method |
|
2 |
Acceptance
criteria |
|
|
3 |
Minimum
five (5) levels of standard solutions |
|
|
4 |
Data
for linear regression equation, Y-intercept, slope, r2 and linearity graph |
|
Range |
1 |
For the assay of drug substances: 80% - 120% of
working concentration (WC) |
|
2 |
For the determination of an impurity: from the
reporting level of an impurity to 120% of the specification |
|
|
Accuracy |
1 |
Testing Method |
|
2 |
Acceptance criteria |
|
|
3 |
Minimum three (3) levels of concentration in
triplicates covering the specified range |
|
|
4 |
Result: reported as percent recovery by the assay
of known added amount of analyte in the sample, OR as the difference between
the mean and the accepted true value together with the confidence intervals |
|
|
Precision
(Repeatability) |
1 |
Testing Method (using sample/product as the test
solution) |
|
2 |
Acceptance criteria |
|
|
3 |
Minimum three (3) levels of concentration in
triplicates covering the specified range, OR minimum six (6) replicates at
100% of the WC |
|
|
4 |
Result: SD, RSD and confidence Interval |
|
|
Precision
(intermediate precision/ ruggedness) |
1 |
Testing Method (using sample/product as the test
solution) |
|
2 |
Acceptance criteria |
|
|
3 |
Minimum three (3) levels of concentration in
triplicates covering the specified range, OR minimum six (6) replicates at 100% of the WC |
|
|
4 |
Cover at least 2 parameters among variation of
analyst, date and equipment |
|
|
5 |
Result: SD, RSD and confidence Interval |
|
Detection
Limit |
1 |
Testing Method: visual
observation/signal-to-noise / standard deviation of the response and the
slope |
|
2 |
If based on standard deviation of the response
and the slope method |
|
|
Minimum five (5) levels of standard solutions |
|
||
Peak area values for all concentrations |
|
||
Data for linear regression equation, Y-intercept,
slope, r2 and linearity graph. |
|
||
3 |
Calculation/formulation where applicable |
|
|
4 |
Related Chromatogram(s) at LOD |
|
|
5 |
Value of detection limit |
|
|
Quantitation
Limit |
1 |
Testing Method: visual
observation/signal-to-noise / standard deviation of the response and the
slope |
|
2 |
if based on the visual observation method,
accuracy and precision data at LOQ must be provided |
|
|
3 |
If based on the calibration curve method |
|
|
Minimum five (5) levels of standard solutions |
|
||
Peak area values for all concentrations |
|
||
Data for linear regression equation, Y-intercept,
slope, r2 and linearity graph. |
|
||
5 |
Calculation/formulation where applicable |
|
|
6 |
Value of quantitation limit |
|
|
System Suitability
Testing |
1 |
RSD, tailing factor, theoretical plate |
|
2 |
Resolution (if two or more components) |
|
|
Robustness
(not mandatory) |
1 |
Testing Method |
|
2 |
Acceptance criteria |
|
|
3 |
Result: refer to acceptance criteria for accuracy
and precision (repeatability) |
|
Checklist for Analytical Method Validation (Chemical)
Dissolution
TEST |
DISSOLUTION |
||
PARAMETER |
No. |
DOCUMENTS REQUIRED |
AVAILABILITY |
Specificity |
1 |
Testing Method |
|
2 |
Acceptance criteria |
|
|
3 |
Chromatogram/spectrum for the following
solutions: |
|
|
Standard |
|
||
Sample |
|
||
Blank/Placebo |
|
||
Spike solution |
|
||
System suitability tests |
|
||
Linearity |
1 |
Testing Method |
|
2 |
Acceptance criteria |
|
|
3 |
Minimum five (5) levels of standard solutions |
|
|
4 |
Data for linear regression equation, Y-intercept,
slope, r2, and linearity graph. |
|
|
Range |
1 |
Dissolution testing: ± 20% over the specified
range Example 1: if the specification is NLT 75% (Q) of
the labeled amount is dissolved in 45 minutes, the validated range would be
60 – 100% of the label claim Example 2: if the specification for a controlled released
product covers a region from 20% after 1 hour, up to 90%, after 24 hours, the
validated range would be 0 – 110% of the label claim |
|
Accuracy |
1 |
Testing Method |
|
2 |
Acceptance criteria |
|
|
3 |
Minimum three (3) levels of concentration in
triplicates covering the specified range |
|
|
4 |
Result: reported as percent recovery by the assay
of known added amount of analyte in the sample, OR as the difference between
the mean and the accepted true value together with the confidence intervals |
|
Precision
(Repeatability) |
1 |
Testing Method (using sample/product as the test
solution) |
|
2 |
Acceptance criteria |
|
|
3 |
Minimum three (3) levels of concentration in
triplicates covering the specified range, OR minimum six (6) replicates at
100% of the WC |
|
|
4 |
Result: SD, RSD and confidence Interval |
|
|
Precision
(intermediate precision/ ruggedness) |
1 |
Testing Method (using sample/product as the test
solution) |
|
2 |
Acceptance criteria |
|
|
3 |
Minimum three (3) levels of concentration in
triplicates covering the specified range, OR minimum six (6) replicates at
100% of the WC |
|
|
4 |
Cover at least 2 parameters among variations of
analyst, date and equipment |
|
|
5 |
Result: SD, RSD and confidence Interval |
|
|
1 |
RSD, tailing factor, theoretical plate |
|
|
2 |
Resolution (if two or more components) |
|
|
Robustness
(Not Mandatory) |
1 |
Testing Method |
|
2 |
Acceptance criteria |
|
|
3 |
Result: refer to acceptance criteria for accuracy
and precision (repeatability) |
|
Commonly Acceptance Criteria
PARAMETER |
ACCEPTANCE
CRITERIA |
Specificity |
Absence of interfering peaks in the placebo,
impurity demonstrates specificity |
Pass peak purity test (particularly for related
substances test) |
|
Linearity |
r2 ≥ 0.995 |
y-intercept at 100% working concentration ≤ 2% |
|
Accuracy |
Measured recovery within 95% - 105% or mean
difference ± 2% & CI |
Precision (Repeatability) |
RSD ≤ 2.0% & CI |
Precision
(intermediate precision/ ruggedness) |
RSD ≤ 2.0% & CI or mean difference ± 2% & CI |
Detection
Limit |
LOD peak must be visible |
If based on a standard deviation of the response
and the slope method, DL = 3.3 Ó/S |
|
If based on signal to noise, S/N= 3:1 or 2:1 |
|
Quantitation
Limit |
if based on the visual observation method,
accuracy and precision data at LOQ must be ± 20% |
If based on a standard deviation of the response
and the slope method, DL = 10 Ó/S |
|
If based on signal to noise, S/N= 3:1 or 2:1 |
|
|
RSD ≤ 2% |
Theoretical plate/column efficiency, N ≥ 2000 |
|
Tailing factor < 2 |
|
Resolution > 2 |
|
Robustness (Not
mandatory) |
Refer to acceptance criteria for accuracy and
precision (repeatability) |