- Identify the source of the leak: The first step is to locate the source of the leak. Look for liquid on the exterior of the HPLC system as well as wet patches on the floor or bench. Examine all of the fittings, connectors, and tubing for evidence of wear, cracks, or corrosion. Inspect the system for leaks using a flashlight or a UV lamp. Leaks can be audible or detected by the presence of bubbles in the system in some instances.
- Tighten the fittings: If the leak is coming from a fitting, use the proper wrench or pliers to tighten it. Overtightening might harm the fitting or cause it to strip, so be careful. After tightening, inspect the fitting for leaks once more.
- Replace the tubing: If the leak is caused by damaged or corroded tubing, replace it. Use tubing that is compliant with the HPLC system's liquid and pressure requirements. To guarantee a tight seal, cut the tubing cleanly and squarely.
- Examine the seals: If the leak is caused by the seals, inspect them for wear and tear or damage. If required, replace the seals with new ones. Before replacing the seals, be sure you grease them with the proper oil.
- If the leak is coming from the column, remove it from the system and check it for signs of damage or wear and tear. If required, replace the column with a new one.
- Check the pump: If the leak is coming from the pump, look for evidence of wear and tear or damage to the pump head. If required, replace the pump head with a new one.
- Check the solvent reservoir: If the leak is coming from the solvent reservoir, look for evidence of wear and tear or damage on the cap. If required, replace the cap with a new one.
- Test the system: After fixing the leak, test the HPLC system by running a sample and checking for any changes in the peak shape or retention time. If the results are consistent and reproducible, the leak has been fixed.
Possible
Cause |
Prevention/Solution |
Loose fitting |
Lighten |
Stripped fitting |
Replace |
Overtightened' fitting |
Loosen and retighten Replace |
Dirty fitting |
Disassemble and clean Replace |
Mismatched parts |
Use all parts from same brand |
Possible
Cause |
Prevention/Solution |
loose check valves |
Tighten check
valve (do not overtighten) Replace check valve |
loose fittings |
Tighten fittings (do not overtighten) |
Mixer seal failure |
Replace mixer seal Replace mixer |
Pump seal failure |
Repair or replace |
Pressure transducer failure |
Repair or replace |
Pulse damper failure |
Replace pulse damper |
Proportioning valve failure |
Check diaphragms, replace if leaky Check for fitting
damage, replace |
Purge valve |
Tighten valve Replace purge
valve |
Possible
Cause |
Prevention/Solution |
Rotor seal failure |
Rebuild or replace injector |
Blocked loop |
Replace loop |
Loose injection-port seal |
Adjust |
Improper syringe-needle diameter |
Use correct syringe |
Waste-line siphoning |
Keep waste line above surface waste |
Waste-line blockage |
Replace waste line |
Possible
Cause |
Prevention/Solution |
Loose end-fitting |
Lighten end-fitting |
Column packing in ferrule |
Disassemble, rinse
ferrule, reassemble |
Improper frit thickness |
Use proper frit (see chart below) |
Possible
Cause |
Prevention/Solution |
Cell gasket failure |
Prevent excessive
backpressure Replace gasket |
Cracked cell window(s) |
Replace window(s) |
Leaky fittings |
Tighten or replace |
Blocked waste line |
Replace waste line |
Blocked now cell |
Rebuild or replace |