Troubleshooting for Increased Pressure

Repeated analysis may increase back pressure. Continuous use of HPLC columns under high pressure can cause deterioration and overload of the equipment. Therefore, it is important to monitor column back pressure regularly and solve the problem timely.

Identification of the Clogging Site
  • The back pressure increase can be due to the clogging of a column or clogging of the equipment. First of all, identify the clogging site.
  • Disconnect the system components one by one to identify the clogged component(s). Start by disconnecting the column from the system and measuring the flowing mobile phase's pressure. The pressure should be close to zero. If the flow pressure without a column is normal, then the pressure increase is due to the clogged column. 
  • The cause of clogging needs to be determined and preventative measures implemented. The column may need to be replaced.

Clogging of Equipment

Identify the specific clogging site according to the method above.


Case 1: High pressure caused by clogged tubing

Cause

Solution

Salt deposit in the tubing.

Disconnect the column and any other equipment before pumping water through the tubing. Washing in a reversing direction is also an effective way. If the situation does not improve, replace the tubing with a new one.


Case 2: High pressure caused by clogged pump

Cause

Solution

The line filter of pump is clogged.

Take apart the line filter, and soak it in the solvent, then clean in an ultrasonic cleaner. If the situation does not improve, replace the line filter with a new one.


Case 3: High pressure caused by clogged manual injector

Cause

Solution

The manual injector is clogged.

Inject with 20ml of contaminant-dissolving solvent (e.g., methanol) by syringe. Wash both lines in the LOAD and INJECT positions. Cleaning the injector in an ultrasonic bath is also effective. If solids caused clogging, wash the injector in a reversing direction. If the situation does not improve, replace the injector with a new one.



What should we do when a clogged column caused pressure increase?

Case 1: Salt deposit in a column caused by pumping high-organic solvent after using a buffer solution

Cause

Solution

Salt deposit in a column.

Wash columns for 30 minutes at half the normal flow rate using 10% organic solvent (methanol or acetonitrile) in water to dissolve the salt deposits. If the situation does not improve, wash the column with 100% water under the same condition.


Prevention: To switch to high organic solvent concentration after using a buffer, first wash a column with a salt-free mobile phase (with the same concentration of organic solvent as the buffer), then switch to the mobile phase of higher organic concentration.


Example: To change mobile phase from 10/90 (v/v) acetonitrile/20mmol/l phosphate buffer (pH2.5) to 90/10 (v/v) acetonitrile/water, first wash the column for 15 minutes with 10/90 (v/v) acetonitrile/water, and then switch to 90/10 (v/v) acetonitrile/water.

Case 2: The sample is not completely dissolved or unfiltered

Cause

Solution

Column frit is clogged by insoluble samples or impurities.

Connect the column in the reverse direction and disconnect it from the detector, and then wash the column for 30 minutes at half of the usual flow rate with the same mobile phase used for analysis. If the situation does not improve, change the frit in the front end of the column (We can replace end fittings with a paid service fee).


Prevention: It is strongly recommended to filter the sample and/or mobile phase.

Attention: If the column is continually connecting in the reverse direction, it may deteriorate.

Case 3: Protein samples that adsorb easily to the column or samples that are slightly soluble in the mobile phase

Cause

Solution

Samples have been adsorbed to packing material or deposited in a column.

Wash the column for 30 minutes with half of the normal flow rate using a solvent that can dissolve the adsorbed substances. Here are the washing procedures for each column type.

Reversed-phase columns

a.  When absorbed substances are not proteins, wash them with methanol or tetrahydrofuran.

b.   When absorbed substances are proteins, wash with 50-70% of acetonitrile/water (containing 0.1% of trifluoroacetic acid). However, proteins may precipitate in a high concentration of organic solvent.

COSMOSIL SL-II

Wash with methanol, tetrahydrofuran or ethanol.

Fullerene columns

Wash with o-dichlorobenzene, 1,2,4-trichlorobenzene.

COSMOSIL Sugar-D/NH2/HILIC columns

Wash with 50/50 (v/v) acetonitrile/water for NH2-MS and 100% water for Sugar-D and HILIC columns.



Prevention: 
  1. Choose appropriate pretreatment for each sample. 
  2. use guard columns.

Attention: 
  • When washing columns, do not connect the column outflow end to the detector. Let the solvent flow into waste.
  • Excessive washing may deteriorate the performance of columns.
  • Do not use a strong alkaline solution (more than pH 7.5) or a strongly acidic solution (less than pH 1.5) for silica-based packing material.
  • Store columns in manufacturer-recommended storage solvent after washing.
  • If the column performance does not improve after washing, replace the column.

Case 4: Graduate pressure increase over time

Cause

Prevention

Contamination of column due to normal long-term use.

Wash the column.

Column damage due to normal long-term use.

Replace the columns.


No improvement in performance after washing.
When the column performance has not improved after washing, it is recommended to replace the column to lessen the pressure burden on the instrument. You could continue to use the column if the peak shape is acceptable and the maximum pressure is less than 20 MPa.