- Impurities in instrument
- Impurities in sample
- Impurities in mobile phase
- The most common reason for ghost peaks originating in the instrument is carryover attributable to the autosampler (or manual injector). When the needle is dipped into the vial to aspirate the sample, substances in the sample, which become the source of carryover, are adsorbed to the inside and outside surfaces of the needle.
- Those substances which are not eliminated even after needle rinsing are carried over to the next analysis, and appear as ghost peaks. Since substances that exhibit particularly strong adsorption will appear even after several blank analyses (analysis in which only the mobile phase is injected), it may be difficult to identify the autosampler as the source of the ghost peaks
- On the other hand, ghost peaks may also be attributed to instrument contamination.
- When there is an impurity in the sample having the same absorption wavelength as the compound of interest, it results in a ghost peak. Since a ghost peak that originates in the sample does not appear in a blank injection, determining the source is relatively easy.
- The remedy is to devise a sample preparation procedure to eliminate the impurity, such as an extraction or use of a disposable filter, etc.
- Further, when the appearance of a ghost peak is due to the degradation of the sample, that degradation can be inhibited by using an autosampler with a vial cooling feature.
- On the other hand, the source of the ghost peak may not be in the sample itself, but rather, from dissolved oxygen in the sample solvent. This is especially noticeable when using a UV detector.
- Ghost peaks can also be generated from the mobile phase in various ways, including the following.
- Generation of organic substances in the mobile phase due to prolonged use, or dissolution of organic substances from the air in the mobile phase.
- Mobile phase vial contamination due to topping off the existing mobile phase with new mobile phase over a long period of time instead of preparing a fresh bottle each day or for each set of samples.
- Use of contaminated organic solvent and/or water to prepare mobile phase
- Use of contaminated mobile phase reagent
- Due to the high consumption of the mobile phase and the relatively long periods of time it is used before being replaced, pretreatment measures such as filtering that are typically used for samples are not practical for eliminating contaminants in the mobile phase.
- Mobile phase contaminants adhere to the inside of the instrument and tubing, small amounts are inevitably transported to the column along with the mobile phase during the solvent delivery process and are likely to be detected as ghost peaks.
- In reversed-phase analysis using gradient elution, such substances that cause ghost peaks are retained in the column at the start of analysis, and are then eluted as the organic solvent ratio increases, making separation very difficult when they exhibit behavior similar to that of the compound of interest.
- Thus, it is difficult to specify the cause when ghost peaks are of mobile phase origin, and furthermore, it is difficult to address the problem by making changes to the method.
Possible Cause |
Solution |
Contamination |
Flush column to remove contamination; use
HPLC-grade solvent |
Elution of
analytes retained from previous injection |
Flush column with
strong solvent at end of run; end gradient at higher solvent concentration |
Ion-pair chromatography - upset equilibrium |
Prepare sample in mobile phase; reduce injection
volume |
Oxidation of
trifluoroacetic acid in peptide mapping |
Prepare
trifluoroacetic acid solutions fresh daily; use antioxidant |
Reversed-phase chromatography - contaminated
water |
Check suitability of water by running different
amounts through column and measure peak height of interferences as function
of enrichment time; clean water by running it through old reversed-phase
column; use HPLC-grade water |
Reversed-phase
chromatography - contaminated water |
Check suitability of
water by running different amounts through column and measure peak height of
interferences as function of enrichment time; clean water by running it
through old reversed-phase column; use HPLC-grade water |
Unknown interferences in sample |
Use sample cleanup or pre-fractionation before
injection |