Common Errors in Method Validation and How to Prevent Them

Method validation is a critical component of analytical chemistry, ensuring that a method is suitable for its intended purpose by providing reliable and consistent results. Whether in pharmaceutical research, environmental testing, or quality control in manufacturing, validated methods form the backbone of trustworthy data and regulatory compliance. Despite its importance, method validation is fraught with potential pitfalls. Common errors during this process can compromise the accuracy, reliability, and reproducibility of analytical results, leading to incorrect conclusions and potentially costly consequences. This guide explores the most frequent errors encountered in method validation and offers practical strategies to prevent them, helping to ensure robust and dependable analytical methods.


Specificity
  • The validation of analytical methods is undoubtedly a difficult and complex task. Unfortunately this means that mistakes are all too common.
  • In this series of articles are reported some examples for discussion related to the method performance characteristics as listed in the current ICH guidance, ICH Q2(R1), namely: Specificity; Robustness; Accuracy; Precision; Linearity; Range; Quantitation limit; and Detection limit.
  • In this first instalment we will consider some mistakes associated with ‘Specificity’. This characteristic is evaluated for both qualitative and quantitative methods but the aim is different for each. For qualitative
  • methods, the aim is to demonstrate that the method can provide the correct information, e.g., an identification method. For quantitative methods, the aim is to demonstrate that the final result generated by the method is not affected by all the potential interferences associated with the method.
  1. Not setting appropriate acceptance criteria
  2. Not investigating all the potential interferences
  3. Not considering potential changes that could occur in the sample/method being tested

Mistake 1: Not setting appropriate acceptance criteria
  • When the results of a validation study don’t comply with the acceptance criteria defined in the protocol, then either the method is not suitable for its intended use, or the acceptance criteria set in the protocol were inappropriate. I am often asked for help on how to explain why it’s okay that results did not meet the acceptance criteria, and not just for specificity.
  • The usual reason for this problem is that generic acceptance criteria were used, typically predefined in an SOP, and no evaluation of their suitability to the method being validated was performed.

Example 1:
  • An identification method by FTIR, which was based on a percentage match with standards spectra in a database, was being validated.
  • The validation failed because the acceptance criteria for the percentage match was set at 98% and the match in the validation study was always in the region of 97%.
  • On investigation it was determined that the percentage match of 98% had no scientific justification, it was just what had been used before.
  • No investigation of the method had been performed prior to the validation.


Example 2:
  • A chromatographic impurities method was being validated. The method validation SOP defined that impurity peaks should have a resolution of 1.5 and thus an acceptance criterion of 1.5 was set in the validation protocol. During the validation study, one of the impurity peaks had a resolution of 1.4.
  • On review of the method development information, it was found that the resolution of this peak was always around 1.4 and the chromatography had been considered acceptable but this information had not made it into the validation protocol.

TIP: Review all the acceptance criteria defined in the validation protocol against what is known about the method. Assess whether the criteria are reasonable, in terms of the method capability and what is considered acceptable. The use of generic acceptance criteria can be a very useful strategy as long they are used in a scientific manner by assessing what is known about the actual method being validated.


Mistake 2: Not investigating all the potential interferences
  • In order to demonstrate that the final result generated by the method is not affected by potential interferences, it is essential that all the potential interferences are considered. This can sometimes be difficult for complex sample matrices so it is important to identify the constituents of the sample matrix as fully as possible.
  • Additionally, it is easy to overlook other sources of interferences that may be introduced as part of the method such as solvents, buffers, derivatization reagents, etc.

TIP: Carry out a thorough review of all potential interferences when designing the validation protocol, particularly if the sample matrix is complex in nature, or if the sample preparation involves the use of multiple reagents.


Mistake 3: Not considering potential changes that could occur in the sample being tested
  • The potential interferences that are present in a sample matrix can change due to changes in the sample composition. The most common example of this situation is probably sample degradation.
  • In situations where a method will be used for samples of different ages, such as in a stability programme, it is essential that this is taken into account during validation and that it is demonstrated that the method can be used for any sample which may require analysis.
  • This means that for some methods, particularly those which are considered to be stability indicating, the specificity section of the validation protocol should include experiments to gather evidence to prove that the method may be successfully used for stability analysis.
  • For methods which analyse the degradation products it would be expected that forced degradation studies were performed during method development to allow the creation of a method that can separate all the components of interest. For other methods this may not have been necessary in method development but a forced degradation study may now be required as part of method validation to demonstrate that the method is stability indicating.

TIP: Consider the long term use of a method when designing the validation protocol. What samples will be tested and are there any anticipated changes that could occur to the samples that would affect the potential interferences for the method? If the method is to be used for stability testing, are there any additional requirements, such as a degradation study?

Robustness
The common mistakes that I have selected for discussion are:
  1. Investigating robustness during method validation
  2. Not investigating the right robustness factors
  3. Not doing anything with the robustness results

  • The purpose of a robustness study is to find out as much as possible about potential issues with a new analytical method and thus how it will perform in routine use. Usually, we deliberately make changes in the method parameters to see if the method can still generate valid data. If it can, it implies that in routine use small variations will not cause problems. This definition is provided in the ICH guideline: “
  • The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage.
  • There is another aspect to robustness that doesn’t neatly fit under this definition which applies to the performance of consumable items in the method, such as chromatography columns. The performance of the column when different batches of the same column packing are used may vary.
  • Although column manufacturers aim for batch to batch reproducibility, most practitioners of HPLC will have come across at least one example of this problem. Another issue is the aging of the column, the column performance generally decreases with age and at some stage the column will have to be discarded. Strictly speaking, these column challenges would actually come under the heading of intermediate precision, following the ICH guideline, but it makes much more sense to investigate them during method development as part of robustness.
  • The method validation guidelines from both ICH and FDA mention the importance of robustness in method development and how it is a method development activity but they do not define whether it needs to be performed under a protocol with predefined acceptance criteria. Since the use of a protocol is a typical approach in most pharma companies it brings me to my first common mistake associated with robustness.

Mistake 1: Investigating robustness during method validation
  • What I mean by this is that the robustness investigation is performed during the method validation, i.e. the outcome of the investigation is not known. I do not mean the approach where the robustness has already been fully investigated and then it is included as a section in the validation protocol for the sole purpose of generating evidence which can be included in the validation report.
  • If robustness is investigated during validation for the first time, the risk is that the method may not be robust. Any modifications to improve robustness may invalidate other validation experiments since they are no longer representative of the final method. It will of course depend on what modifications have to be made.
  • As FDA suggests… “During early stages of method development, the robustness of methods should be evaluated because this characteristic can help you decide which method you will submit for approval.”


TIP: If for some reason robustness hasn’t been thoroughly evaluated in method development then investigate it prior to execution of the validation protocol using a specific robustness protocol. If any robustness issues are identified, these can be resolved prior to the validation. The nature of the robustness problems will determine whether the resolution is just a more careful use of words in the written method or if method parameters need to be updated.

Mistake 2: Not investigating the right robustness factors
  • If you choose the wrong factors you may conclude that the method is robust when it isn’t. Typically what happens then is that there are a lot of unexpected problems when the method is transferred to another laboratory, and since transfer is a very common occurrence in pharma, this can be very expensive to resolve.
  • When choosing robustness factors it is tempting to read through the method and select all the numerical parameters associated with instrumentation. For example, when assessing HPLC methods there is a tendency to only look at the parameters of the instrument without consideration of the other parts of the method, such as the sample preparation. Unfortunately, sample preparation is an area where robustness problems often occur. Detailed knowledge of how the method works is required to identify the most probable robustness factors.

TIP:
The most important factors for robustness are often those which were adjusted in method development. Review all the steps in the method to choose robustness factors and use a subject matter expert to help if
necessary.

Mistake 3: Not doing anything with the robustness results
  • The reason for investigating robustness is to gain knowledge about the method and to ensure that it can be kept under control during routine use. Very often robustness data is presented without any comments in the validation report and is not shared with the analysts using the method.
  • This tick-box approach may be in compliance with regulatory guidance but it is not making the most of the scientific data available. The discussion of the method robustness in the validation report should be a very useful resource when the method needs to be transferred to another laboratory and will assist in the risk assessment for the transfer.

TIP:
Review the robustness data thoroughly when it is available and ensure that there is a meaningful discussion of its significance in the validation report.

Accuracy
The common mistakes that I have selected for discussion are:
  1. Not evaluating accuracy in the presence of the sample matrix components
  2. Performing replicate measurements instead of replicate preparations
  3. Setting inappropriate acceptance criteria


  • The definition of accuracy given in the ICH guideline is as follows: ‘The accuracy of an analytical procedure expresses the closeness of agreement between the value which is accepted either as a conventional true value or an accepted reference value and the value found.’ This closeness of agreement is determined in accuracy experiments and expressed as a difference, referred to as the bias of the method.
  • The acceptance criterion for accuracy defines how big you are going to let the bias be and still consider the method suitable for its intended purpose.
  • The term accuracy has also been defined by ISO to be a combination of systematic errors (bias) and random errors (precision) and there is a note about this in the USP method validation chapter, <1225>: ‘A note on terminology: The definition of accuracy in <1225> and ICH Q2 corresponds to unbiasedness only.
  • In the International vocabulary of Metrology (VIM) and documents of the International Organization for Standardization (ISO), accuracy” has a different meaning. In ISO, accuracy combines the concepts of unbiasedness (termed “trueness”) and precision.’
  • From the point of view of performing validation, the difference in the definitions doesn’t make a lot of difference, we usually calculate both bias and precision from the experimental data generated in accuracy experiments.


Mistake 1: Not evaluating accuracy in the presence of the sample matrix components
Since the purpose of the accuracy experiments is to evaluate the bias of the method, the experiments that are performed need to include all the potential sources of that bias. This means that the samples which are
prepared should be as close as possible to the real thing. If the sample matrix prepared for the accuracy experiments is not representative of the real sample matrix then a source of bias can easily be missed or underestimated.

TIP: The samples created for accuracy experiments should be made to be as close as possible to the samples which will be tested by the method. Ideally these ‘pseudo-samples’ will be identical to real samples except that the amount of the component of interest (the true value) is known. This can be very difficult for some types of sample matrix, particularly solids where the component of interest is present at low amounts (e.g., impurities determination).
For impurities analysis, it may be necessary to prepare the accuracy samples by using spiking solutions to introduce known amounts of material into the sample matrix. Although this carries the risk of ignoring the potential bias resulting from the extraction of the impurity present as a solid into a solution, there isn’t really a workable alternative.

Mistake 2: Performing replicate measurements instead of replicate preparations
  • Performing replicate preparations of accuracy ‘pseudo-samples’ allows a better evaluation of what differences in the data are due to the bias and what are due to variability of the method, the precision. A minimum of 9 replicates are advised by the ICH guidance and these should be separate preparations. For solids, this could be 9 separate weighings into 9 separate volumetric flasks, as per the method.
  • However, the preparation does depend on the nature of the sample matrix and the practicality of controlling the known value for the component of interest. As discussed above, sometimes in the case of impurities methods, solutions may be required for practical reasons even though the sample matrix exists as a solid.
  • In this case 9 separate weighings does not offer more representative ‘pseudo-samples’ and thus a single stock solution for the impurity would probably be a better choice.

TIP: Assess the sample matrix and try to prepare separate replicates when
possible so that the data produced is as representative as possible and includes typical sources of variability.

Mistake 3: Setting inappropriate acceptance criteria
  • As mentioned previously, the acceptance criterion for accuracy is based on how much bias you will allow in the results from the method. It is obviously better not to have any bias in a method but there is always a certain amount of potential bias associated with the combination of the sample matrix, the level of the components of interest in the sample, and the instrumentation used for the measurement.
  • For the method to be capable the bias needs to be less than the specification for the result. For example, if a drug substance specification requires that there must be between 99 and 101 %w/w of the drug present, then a method which has a bias of 2% is not going to be acceptable.

TIP: Make sure that the acceptance criteria set for accuracy in method validation are compatible with the requirements for the method, and in particular, the specification for the test.

Post a Comment

Previous Post Next Post
close