- Objective
- Purpose
- Control panel testing
- Standard Operating Procedure verification
- Instrumentation Calibration
- Summary of Operational Qualification
- Verifies that the Instrument consistently operates within a specified set of parameters under normal operational conditions.
- Assure that the Instrument performance is adequate to support the process for which the Instrument is intended for
- Demonstrates the completeness and adequacy of Standard Operating procedure and calibration procedure.
- Check all the components on the key panel are properly labeled/identified
- Turn on the power to the key panel
- Set the required controls
- Verify the function of each Key/component on the panel against its specification
- Observe and record the responses of each component of the panel
- To check the role of the control panel key
Control Panel Key | Use | Observation |
Power Key | To switch on the instrument and switch to standby mode |
|
Power Indicator | Standby- Red On- Green |
|
Arrow Key | Forward-back key for negotiating through menus |
|
Alpha-numeric keys | Keys for entering letters and numbers |
|
Soft keys | To change the function of the display |
|
Checked By:__________ Verified By:__________
- To check the operation of measurement parameters in the measurement screen
Step No | Parameters | Method of Testing | Acceptance Criteria | Observation |
1 | Operator | Enter the name of the operator | The operator's name shall enter |
|
2 | Sample Name | Enter the name of the sample | Sample name shall enter |
|
3 | Measurement Mode | Set optical rotation mode | Set mode shall display as an optical rotation |
|
Set specific optical rotation mode | Set mode shall display a specific optical rotation |
| ||
Set Concentration mode | Set mode shall display as Concentration |
| ||
4 | Cell length | Enter the length of the cell as X mm | Cell length shall show as X mm |
|
5 | Temperature | Select the temperature where the temperature reading required | The display shall show the temperature |
|
6 | Correction factor | Enter correction factor | Automatically correction shall take place |
|
Checked By:__________ Verified By:___________
4. STANDARD OPERATING INSTRUCTIONS VERIFICATION
Sr. No. | SOP TITLE | SOP No. | VERIFIED BY |
1 | Operation and calibration Procedure of POLARIMETER, MODEL- ADP440+ |
|
|
Description | Yes/No | Checked by |
All acceptance Criterion outlined in the Operational qualification was met |
|
|
All applicable data sheets were signed, checked, and verified. |
|
|
Standard operating Instructions/ calibration Procedures were verified for completeness and accuracy |
|
|
| Signatories | Signature | Date |
Executed by | Quality control |
|
|
Reviewed by | Quality control |
|
|
Approved by | Quality Assurance |
|
|