Operational Qualification of Polarimeter

TABLE OF CONTENT
  1. Objective
  2. Purpose
  3. Control panel testing
  4. Standard Operating Procedure verification
  5. Instrumentation Calibration
  6. Summary of Operational Qualification

1. OBJECTIVE
The purpose of this study is to qualify and certify the operational performance of POLARIMETER, MODEL- ADP440+ at PharmaGuide Limited.

This study verifies:
  • Verifies that the Instrument consistently operates within a specified set of parameters under normal operational conditions.
  • Assure that the Instrument performance is adequate to support the process for which the Instrument is intended for
  • Demonstrates the completeness and adequacy of Standard Operating procedure and calibration procedure.

2. PURPOSE
The purpose of this protocol is to establish documentary evidence to ensure that the installed POLARIMETER, MODEL- ADP440+ will operate reproducibly and consistently within its full dynamic range of operation according to manufacturer’s specifications.

3. CONTROL PANEL TESTING

Objective:
To verify that all components on the control panel function as designed.


Testing procedure:
  • Check all the components on the key panel are properly labeled/identified
  • Turn on the power to the key panel
  • Set the required controls
  • Verify the function of each Key/component on the panel against its specification
  • Observe and record the responses of each component of the panel

Control Panel Testing:
  • To check the role of the control panel key

Control Panel Key

Use

Observation

Power Key

To switch on the instrument and switch to standby mode

 

Power Indicator

Standby- Red

On- Green

 

Arrow Key

Forward-back key for negotiating through menus

 

Alpha-numeric keys

Keys for entering letters and numbers

 

Soft keys

To change the function of the display

 


Checked By:__________                         Verified By
:__________


ALSO READ: Operational Qualification of Disintegration Tester

  • To check the operation of measurement parameters in the measurement screen

Step No

Parameters

Method of Testing

Acceptance Criteria

Observation

1

Operator

Enter the name of the operator

The operator's name shall enter

 

2

Sample Name

Enter the name of the sample

Sample name shall enter

 

3

Measurement Mode

Set optical rotation mode

Set mode shall display as an optical rotation

 

Set specific optical rotation mode

Set mode shall display a specific optical rotation

 

Set Concentration mode

Set mode shall display as Concentration

 

4

Cell length

Enter the length of the cell as X mm

Cell length shall show as X mm

 

5

Temperature

Select the temperature where the temperature reading required

The display shall show the temperature

 

6

Correction factor

Enter correction factor

Automatically correction shall take place

 


Checked By:__________                      Verified By:___________

4. STANDARD OPERATING INSTRUCTIONS VERIFICATION
The following Standard Operating Procedures were verified as important for Effective Performance of POLARIMETER, MODEL- ADP440+


Sr. No.

SOP TITLE

SOP No.

VERIFIED BY

1

Operation and calibration Procedure of POLARIMETER, MODEL- ADP440+

 

 


5. INSTRUMENT CALIBRATION VERIFICATION

SOP Title: OPERATION AND CALIBRATION PROCEDURE OF POLARIMETER, MODEL- ADP440+
SOP No:
Calibration Date: 

6. SUMMARY OF THE OPERATION QUALIFICATION

Description

Yes/No

Checked by

All acceptance Criterion outlined in the Operational qualification was met

 

 

All applicable data sheets were signed, checked, and verified.

 

 

Standard operating Instructions/ calibration Procedures were verified for completeness and accuracy

 

 


POST-APPROVAL

 

Signatories

Signature

Date

Executed by

Quality control

 

 

Reviewed by

Quality control

 

 

Approved by

Quality Assurance

 

 