What is distortion for Peaks in the Chromatogram?

Chromatography is a common analytical method used to separate and identify chemical components in a mixture. A mixture is passed between a stationary phase and a mobile phase, which separates the components depending on their physical and chemical characteristics. A chromatogram, which is a graphical depiction of the chemical components in the combination, is the outcome.

Peak distortion is a typical problem that occurs during the chromatography process. Peak distortion refers to variations in the form, size, and placement of chromatogram peaks. This can occur for a variety of causes, including the presence of contaminants, changes in temperature or pressure, changes in the mobile or stationary phase, and so on.


The presence of peak distortion in the chromatogram can lead to inaccurate or unreliable results. It can also affect the sensitivity and selectivity of the analytical method. Therefore, it is essential to identify and eliminate peak distortion to ensure the accuracy and precision of the analytical results.


There are several types of peak distortions that can occur in the chromatogram. One of the most common types is fronting or tailing. Fronting occurs when the peak appears broader at the leading edge, whereas tailing occurs when the peak appears broader at the trailing edge. This type of distortion can occur due to several factors, such as column overload, poor column packing, sample overload, or column contamination.

Splitting is another kind of peak distortion that happens when a single peak appears as many peaks. This form of distortion might occur as a result of the presence of closely related chemicals, column overloading, or incorrect detector response.

Peak overlap is another sort of distortion that happens when two or more peaks in a chromatogram overlap. This may occur as a result of the existence of identical substances or column overloading.


What is Distortion for Peaks in the Chromatogram?
Distortion of peaks in the chromatogram usually is the result of a physical, rather than a chemical problem. When all peaks in the chromatogram show the same type of distortion, the initial problem occurred before the analytes began migrating through the column.


A common cause of this type of peak distortion is a partially blocked frit or a void at the head of the column. Columns are not very susceptible to voids unless they are operated outside of their recommended pH range.  Frit blockage remains a common problem.  

The frit at the column inlet typically is 2.0 μm porosity for 5 μm particle columns and 0.5 μm porosity for columns ≤ 3 Î¼m. If particulate matter from the sample, worn pump seals, or the mobile phase reaches the column, it usually collects on the frit. 


This material can distort the distribution of the sample at the column inlet, that part of the sample reaches the column via a different flow path and thus later than another portion of the sample. Since no separation has taken place at this point, all analytes are distorted in the same manner and the chromatogram shows distortion for all peaks.

To prevent this problem, filter the mobile phase if it has the potential to contain particles (e.g. buffer precipitate or dust); replace the pump seals before they wear enough to shed particles. If the sample contains particulate debris, either filter it or centrifuge it briefly to remove particles. 
To avoid this it is strongly recommended to install a 0.5 μm porosity in-line filter just downstream from the autosampler to catch any particulate matter that inadvertently enters the HPLC system. The system backpressure will rise when this in-line filter frit becomes blocked; replacement of the frit is a simple, fast, and inexpensive task. The use of an in-line filter is one of the least expensive ways to prolong the column lifetime.


Conclusion
Several solutions may be used to avoid peak distortion in chromatograms. One of the most popular ways is to optimize the chromatographic conditions, such as temperature, pressure, mobile phase composition, and flow rate. Furthermore, effective sample preparation, such as filtering, dilution, or derivatization, can aid in the elimination of peak distortion. Peak distortion may also be avoided by performing routine maintenance and cleaning on chromatographic equipment such as the column, detector, and injector.

In conclusion, peak distortion is a common issue that can occur during the chromatography process. It can affect the accuracy and precision of the analytical results and should be identified and eliminated to ensure reliable and trustworthy analytical data. Employing appropriate chromatographic conditions, sample preparation, and equipment maintenance can help to minimize peak distortion in the chromatogram.