- OOT results are test results that deviate significantly from historical data trends, even though they are not necessarily out of specification (OOS). This can occur in stability studies, manufacturing processes, or quality control tests.
- An OOT event is typically triggered when a result, though within specification, shows a shift from past data or established trends. For instance, if stability data for a batch of drugs shows a gradual potency loss over time, but one result falls significantly lower than expected at a particular time point, it would be considered OOT.
- Investigating OOT involves performing a detailed root cause analysis. The aim is to determine why the result deviates from the expected trend. This could be due to factors such as:
- Variability in raw material quality
- Manufacturing process deviations
- Analytical method issues
- Environmental or storage conditions impacting product stability
- Once the cause is identified, a risk assessment is conducted to evaluate the impact on product quality, patient safety, and regulatory compliance. If necessary, corrective actions are implemented, such as changes to the process or further testing.
- If the OOT investigation identifies a specific problem, corrective and preventive actions (CAPA) are implemented to avoid future occurrences. This may involve adjusting the manufacturing process, retraining staff, or improving raw material sourcing.
Date of OOT occurrence: |
OOT Reference No.: |
|||||||||||||
A] OOT REPORTING: (To be completed by the analyst) |
||||||||||||||
Product / Material Name |
|
QC Reference No. |
|
|||||||||||
Batch No. |
|
Mfg. Date |
|
|||||||||||
Specification No. |
|
Expiry Date |
|
|||||||||||
Test Name |
|
Test Method No. |
|
|||||||||||
Stage of testing (For Drug substance / Intermediate /
Drug product) |
In-process Finished product / Intermediate release testing Stability Months Accelerated / Long Term / Intermediate |
|||||||||||||
Summary of OOT Test Results (state result and specification) |
||||||||||||||
|
||||||||||||||
Details of abnormal observations noted during the testing if any |
||||||||||||||
|
||||||||||||||
Analyst Name |
Signature & Date |
|
||||||||||||
B] LABORATORY INVESTIGATION: |
||||||||||||||
Sr. No. |
Check Parameters |
Observations (Yes/No/NA) |
Comments |
|||||||||||
1 |
Sampling |
|
|
|||||||||||
|
Any abnormal observation noted during sampling? |
|
|
|||||||||||
Any noticeable difference in sample appearance? |
|
|
||||||||||||
A sample is correctly collected & labeled? |
|
|
||||||||||||
Sample storage was done appropriately? |
|
|
||||||||||||
2 |
General |
|
|
|||||||||||
|
Any unusual happening in lab? (e.g. power failure) |
|
|
|||||||||||
Was the method discussed with the analyst? |
|
|
||||||||||||
The correct analytical method used? |
|
|
||||||||||||
The analyst was trained to perform the test? |
|
|
||||||||||||
Correct glassware used for dilutions? |
|
|
||||||||||||
Glassware were properly cleaned? |
|
|
||||||||||||
Instruments used are qualified? |
|
|
||||||||||||
Instruments used within calibration validity period |
|
|
||||||||||||
Instrument Used (Name & Id) |
Calibration Due |
|
||||||||||||
|
|
|
|
|||||||||||
|
|
|
|
|||||||||||
|
|
|
|
|||||||||||
|
|
|
|
|||||||||||
Instrument setup & operation as per standard operating procedure? |
|
|
||||||||||||
Use of appropriate grade of chemical and reagents within the
validity period? |
|
|
||||||||||||
Correct normality/molarity of volumetric solutions used? |
|
|
||||||||||||
VS used |
Valid up to date |
Strength |
||||||||||||
|
|
|
|
|||||||||||
|
|
|
|
|||||||||||
3 |
Sample / Standards Preparation |
|||||||||||||
|
Sample & Std preparations done as per the test method? |
|
|
|||||||||||
Is any weighing error identified? |
|
|
||||||||||||
Correct potency of the standard
used in the calculation? Standard is within the validity period? |
|
|
||||||||||||
Std(s) Used |
Valid up to date |
Potency |
|
|||||||||||
|
|
|
|
|||||||||||
|
|
|
|
|||||||||||
Is the sample properly shaken, sonicated or heated / warmed
as per method of analysis? |
|
|
||||||||||||
Are the sample / standard dilutions correctly performed as per
method of analysis? |
|
|
||||||||||||
Any noticeable preparation? Difference noted in the sample/Standard |
|
|
||||||||||||
Are the samples filtered/ centrifuged / membrane filtered properly before introduction into the instrument or analysis by classical method? |
|
|
||||||||||||
Are samples/standards preparations stored under correct
environment/time before analysis? |
|
|
||||||||||||
Tablets/granules are ground properly? |
|
|
||||||||||||
Any errors in calculation and transcription? |
|
|
||||||||||||
4 |
Chromatography |
|||||||||||||
|
Correct Column Used (E.g. column
make, Dimension, Particle Size, End capped / Non-End capped, Pore Size,
Carbon Loading)? |
|
|
|||||||||||
Any leakages observed in the fittings? |
|
|
||||||||||||
Correct instrument parameters
used (e.g. for HPLC – type of detector, flow rate, oven temp., wavelength, injection volume, sample temp. For GC – type of detector, flow rate, oven temp.
Injection volume, Inj. Temp, detector temp.)? |
|
|
||||||||||||
Mobile phase preparation is as per the method (check for composition, pH, air bubbles)? |
|
|
||||||||||||
Any unusual or unexpected response observed with standard
or test preparations? |
|
|
||||||||||||
System suitability acceptance criteria were met during the analysis? |
|
|
||||||||||||
5 |
Dissolution |
|||||||||||||
|
Correct instrument parameters used (apparatus type, speed,
bath temperature, time, medium used, volume)? |
|
|
|||||||||||
Dissolution Medium Degassed? |
|
|
||||||||||||
Sample Withdrawn Correctly? |
|
|
||||||||||||
Correct Filter Used? |
|
|
||||||||||||
6 |
Microbiological Assay |
|||||||||||||
|
Was media from a single prepared lot used in the assay? |
|
|
|||||||||||
Was there any malfunction or breakdown of the incubator? |
|
|
||||||||||||
Was the temperature of the
incubator during the incubation period as per requirement? |
|
|
||||||||||||
Are the zones of inhibition/exhibition clearly defined? |
|
|
||||||||||||
Is the merging of zones seen? |
|
|
||||||||||||
Was zone reading done as per SOP? |
|
|
||||||||||||
Is the Zone reader/vernier calipers in the calibrated state? |
|
|
||||||||||||
Instrument ID: Calibration due
date: |
|
|
||||||||||||
7 |
Stability |
|||||||||||||
|
Any malfunction or breakdown of stability chamber? |
|
|
|||||||||||
Any failure of utilities (power, water, UPS)? |
|
|
||||||||||||
Any deviation in temperature / humidity monitoring? |
|
|
||||||||||||
Any damage to pack? |
|
|
||||||||||||
Any deviation from SOP for sample pull out time? |
|
|
||||||||||||
Were samples, after pull out,
stored as per the conditions specified in the SOP? |
|
|
||||||||||||
Were samples analysed within the specified time period as in
the SOP? |
|
|
||||||||||||
Any change in method of analysis or specification? |
|
|
||||||||||||
Any other (to be specified) |
|
|
||||||||||||
8 |
Any other findings (review of method validation data, Trend data etc.): |
|||||||||||||
9 |
Was similar OOT reported for this product earlier? If yes, state the identified cause and corrective / preventive actions taken that time – |
|||||||||||||
10 |
LABORATORY ERROR IDENTIFIED : Yes / No |
|||||||||||||
If yes, describe the error: |
||||||||||||||
QC Analyst (Sign & Date) |
Section Incharge of QC (Sign & Date) |
|||||||||||||
11 |
ACTION TO BE FOLLOWED ( In case of lab error) |
Yes/No |
Comment |
|||||||||||
|
Retesting of retained sample solution by Original Analyst |
|
|
|||||||||||
Retesting of retained sample by Original Analyst |
|
|
||||||||||||
Correction in document |
|
|
||||||||||||
Any other (To be specified) |
|
|
||||||||||||
Section In-charge Of QC (Sign & Date): |
|
|||||||||||||
12 |
ACTION TO BE FOLLOWED (In case of obvious error related to sampling or sample handling) |
|||||||||||||
REASON FOR RE-SAMPLING: |
||||||||||||||
RE-SAMPLING PLAN: |
||||||||||||||
Total number of containers |
|
|||||||||||||
Number of containers to be sampled |
|
|||||||||||||
Sample quantity from each container |
|
|||||||||||||
Remarks or Special instructions: |
||||||||||||||
QC In-charge (Sign & date) |
QA In-charge (Sign & date) |
|||||||||||||
13 |
RESULT OF RETEST (in case of obvious lab / sampling error): |
|||||||||||||
1. |
2. |
Average = |
||||||||||||
QC Analyst (Sign & Date) |
QC In-charge (Sign & Date) |
|||||||||||||
CONCLUSION: OOT test result Valid, Report the initial result
Invalid, Report new results
If OOT is valid, ÿ Initiation of Manufacturing Investigation |
||||||||||||||
QA In-charge: (Sign & Date) |
||||||||||||||
C] |
MANUFACTURING INVESTIGATION: |
|||||||||||||
For details, refer to Annexure 4 - Manufacturing Investigation Form. Manufacturing Error detected: Yes / No |
||||||||||||||
Production In-charge (Sign & Date) |
QA In-charge (Sign & Date) |
|||||||||||||
|
QA decision on disposition of the batch: QA In-charge (Sign & Date) |
|||||||||||||
D] |
CLOSURE OF OOT INVESTIGATION |
|||||||||||||
|
Summary and Conclusion: (Based on report of Laboratory / Manufacturing investigations)
Assignable Cause:
Corrective Action and Preventive Action: If CAPA not completed within 30 days, CAPA ref. no. |
|||||||||||||
QC In-charge (Sign & Date) |
Production In-charge (Sign & Date) |
QA In-charge (Sign & Date) |
||||||||||||
List of Attachments (if any): |