Diagnostics Tips and Tricks for Retention Time in HPLC

What is the retention time in HPLC?
Retention time (RT) is a measure of the time taken for a solute to pass through a chromatography column. It is calculated as the time from injection to detection. The RT for a compound is not fixed as many factors can influence it even if the same HPLC and column are used.


How do you increase retention time in HPLC?
An analyst can change its retention by changing the polarity of the mobile phase (the mobile phase is competing with the solute). Thus a more polar mobile phase will increase retention, and a more non-polar mobile phase will decrease retention.

How do I reduce retention time in HPLC?
As temperature is increased, retention will decrease. If the room experiences wide temperature fluctuations, the HPLC retention times will probably be affected. The best solution is to run analyses at a temperature that can be controlled by using an oven.

What factors affect retention time in HPLC?
For a particular compound, the retention time will vary depending on:
  1. the pressure used (because that affects the flow rate of the solvent)
  2. the nature of the stationary phase (not only what material it is made of, but also particle size)

Retention Time
  1. Retention time changed to a new constant value
  2. Erratic retention time
  3. Increasing retention time
  4. Decreasing retention time


Retention time changed to a new constant value: 
  • Check column, mobile phase, temperature, method, flow rate

Increasing/decreasing retention times: 
  • Column contaminated, degraded
  • System not equilibrated
  • Mobile phase contaminated

Erratic retention times: 
  • Check if the system also has erratic pump pressure  or pressure fluctuations
  • System not equilibrated
  • Check for leaks
  • Improper solvent blending
  • Temperature fluctuations


Diagnostics Tips and Tricks for Retention Time in HPLC

High-performance liquid chromatography (HPLC) is an essential analytical tool in the pharmaceutical, food, and environmental industries. It is used for the separation, identification, and quantification of chemical compounds in complex mixtures. One of the critical parameters in HPLC analysis is the retention time, which is the time it takes for a compound to travel through the column and elute from the detector. Retention time is influenced by several factors, including column type, mobile phase composition, temperature, and flow rate. Accurate retention time is crucial for identifying compounds and ensuring reproducibility. In this article, we will discuss some diagnostics tips and tricks for retention time in HPLC.


  • The column is the heart of the HPLC system and is the primary factor affecting retention time. To ensure accurate retention time, it is essential to maintain the column in good condition. The column should be cleaned and conditioned regularly to remove any contaminants that may affect the separation. 
  • The recommended frequency for column cleaning depends on the type of column and the sample matrix. For example, a reverse-phase column used for analyzing proteins in a complex matrix may need to be cleaned more frequently than a normal-phase column used for analyzing simple compounds.

  • The mobile phase composition is another significant factor affecting retention time. The solvent strength, pH, and buffer concentration can all affect the retention time. It is essential to optimize the mobile phase composition to achieve the desired separation and retention time. If the retention time is too short or too long, the mobile phase composition can be adjusted by changing the solvent strength, pH, or buffer concentration.

3. Temperature Control
  • Temperature is another crucial factor that affects retention time. An increase in temperature generally decreases retention time, while a decrease in temperature increases retention time. Therefore, it is essential to maintain the temperature of the HPLC system and the mobile phase within a narrow range to achieve reproducible retention time. Temperature control is particularly important for thermally sensitive compounds that may degrade or react with the mobile phase or the column material.

4. Flow Rate Control
  • The flow rate is another important parameter that affects retention time. An increase in flow rate generally decreases retention time, while a decrease in flow rate increases retention time. Therefore, it is essential to maintain a consistent flow rate throughout the analysis to achieve reproducible retention time. The flow rate can be controlled by adjusting the pump flow rate or the column backpressure. It is recommended to use a low flow rate (e.g., 0.2-1.0 mL/min) for most HPLC applications to achieve optimal separation and retention time.

5. Detector Calibration
The detector is the final component in the HPLC system and is responsible for detecting the eluting compounds. To ensure accurate retention time, it is essential to calibrate the detector regularly. Detector calibration involves adjusting the detector sensitivity and wavelength accuracy to ensure that the detector response is consistent and reproducible. Detector calibration should be performed before each analysis to ensure accurate retention time.

In conclusion, accurate retention time is crucial for successful HPLC analysis. To achieve accurate retention time, it is essential to maintain the column, optimize the mobile phase composition, control the temperature and flow rate, and calibrate the detector regularly. By following these diagnostics tips and tricks, you can achieve reproducible and reliable results in your HPLC analysis.

Post a Comment

Previous Post Next Post
close